OCT: interpretation and pitfalls

  • Module 2 of 15
  • About 15 minutes
  • Level 2
How to Interpret an OCT Scan in Glaucoma: A Step-by-Step Guide Plays on YouTube. By pressing play you agree to YouTube's cookies.

Animation and script by Ronald de Crom, ophthalmologist and glaucoma specialist. Narration: synthetic voice.

Learning objectives

After this module you can:

  • Interpret the RNFL and macular analyses systematically
  • Recognise artefacts, red disease and green disease
  • Use OCT to follow progression and know its limits

Key points

  • Read systematically: scan quality and segmentation first, then the RNFL (average, sectors, TSNIT profile, deviation map), then the macular ganglion cell analysis. Compare both eyes and check that the findings match the disc and the visual field.

  • Artefacts are common: segmentation errors from vitreous traction, an epiretinal membrane or peripapillary atrophy, a decentred circle, a low signal from cataract or a dry ocular surface, and motion. When a value is unexpected, look at the B-scan itself.

  • Red disease (false positives): high myopia, tilted, small or large discs and peripapillary atrophy shift the nerve fibre bundles, and the normative database may not fit the patient's age or ethnicity. Red flags then appear without glaucoma.

  • Green disease (false negatives): a focal defect can be averaged away in a quadrant value, an eye that started with a thick RNFL can lose a lot and stay green, and an epiretinal membrane or oedema can thicken the layer. A green printout does not rule out glaucoma.

  • The macular ganglion cell analysis shows damage close to fixation, often as an inferior arcuate defect that respects the horizontal midline. It helps in high myopia, where the disc scan is less reliable, but other macular disease confounds it.

  • For progression, compare good-quality scans from the same device with trend or event analysis. Change must exceed the test–retest variability (a few micrometres for the average RNFL). In advanced disease the RNFL reaches a floor of about 50 µm; follow the visual field and the macular scan instead.

Before you call it glaucoma on OCT

  1. Is the scan of good quality, centred and correctly segmented?
  2. Does the patient fit the normative database (refraction, disc size, age)?
  3. Does the pattern fit glaucoma: superior or inferior poles, arcuate on the macula?
  4. Does it match the disc and the visual field?
  5. Is it reproducible on a repeat scan?

Test yourself

Five questions. Choose an answer to see the explanation.

1. A patient with −8 dioptres of myopia has several red sectors, a normal disc and a normal visual field. What is the most likely explanation?
2. The average RNFL suddenly increases by 15 µm in one eye. The B-scan shows an epiretinal membrane. What does this mean?
3. The software reports a very thin RNFL in one clock hour, but the disc looks healthy there. What is the first step?
4. A patient changes clinics and brings OCT results from another brand of device. Can you compare the RNFL values directly?
5. Why is OCT less useful for follow-up in advanced glaucoma?