PreserFlo MicroShunt

A tiny tube that drains fluid to a bleb under the conjunctiva, with a simpler procedure than trabeculectomy.

In short

PreserFlo MicroShunt Explained Plays on YouTube. By pressing play you agree to YouTube's cookies.

Animation and script by Ronald de Crom, ophthalmologist and glaucoma specialist. Narration: synthetic voice.

The PreserFlo MicroShunt is a soft, very thin tube of less than a centimetre. It drains fluid from the front of the eye to a small reservoir (bleb) under the outer layer of the eye, similar to a trabeculectomy.

The operation is shorter, and no flap or opening needs to be made in the wall of the eye.

In depth

Device
A tube made of SIBS (poly(styrene-block-isobutylene-block-styrene)), 8.5 mm long with a 70 µm lumen, implanted ab externo with mitomycin C.
Indication
Mild to advanced open-angle glaucoma not controlled on maximal tolerated medical therapy.
Evidence
In a randomised comparison with trabeculectomy, IOP reduction was somewhat smaller, with fewer postoperative interventions.
Complications
Early hypotony, bleb fibrosis and failure, and (rarely) tube exposure, migration or contact with the cornea or iris.

Surgical technique

PreserFlo MicroShunt – Step by Step Plays on YouTube. By pressing play you agree to YouTube's cookies.

Surgical footage is shown with written patient consent.

Surgical steps

  1. Fornix-based conjunctival and Tenon's dissection
  2. Application of mitomycin C
  3. Mark 3 mm from the limbus and create a short scleral pocket
  4. Needle track into the anterior chamber with a 25-gauge needle
  5. Insert the MicroShunt bevel up; wedge the fin into the scleral pocket
  6. Confirm flow at the distal end and place the tube under Tenon's
  7. Close Tenon's and conjunctiva, watertight

Pearls

  • The tube should lie in the anterior chamber parallel to the iris, clear of the cornea.
  • Confirm spontaneous flow before closure; if absent, prime the tube with a cannula.