Measuring eye pressure

  • Module 4 of 13
  • About 10 minutes
  • Level 1
Eye Pressure Measurement (IOP) Explained Plays on YouTube. By pressing play you agree to YouTube's cookies.

Animation and script by Ronald de Crom, ophthalmologist and glaucoma specialist. Narration: synthetic voice.

Learning objectives

After this module you can:

  • Explain what intraocular pressure (IOP) is and how to interpret the normal range
  • Describe the main tonometry methods and when to use each
  • Recognise why a single reading can be misleading

Key points

  • IOP is measured in mmHg. In the general population the mean is about 15–16 mmHg, and values above 21 mmHg are statistically high. Glaucoma also occurs at normal pressures, and many people with a high pressure never develop it.

  • Goldmann applanation tonometry (GAT) is the reference standard. It measures the force needed to flatten a corneal area 3.06 mm in diameter, after a topical anaesthetic and fluorescein.

  • Rebound tonometry (such as iCare) needs no anaesthetic and is useful in children and for home monitoring. Air-puff tonometry is mainly used for screening.

  • Central corneal thickness affects the reading: a thin cornea gives a falsely low value, a thick cornea a falsely high one. Refractive laser surgery also lowers the measured value.

    Watch also

    Pachymetry

    How corneal thickness is measured, and why it matters for the pressure reading.

    Pachymetry Explained Plays on YouTube. By pressing play you agree to YouTube's cookies.
  • No correction formula for corneal thickness is reliable enough for routine use. Use corneal thickness as context and as a risk factor, not to recalculate the pressure.

  • IOP varies during the day and is often highest in the early morning. One office reading can miss the peak; a diurnal curve (phasing) can help.

Good practice when measuring with Goldmann

  1. Check the calibration regularly and disinfect the tonometer tip between patients.
  2. Hold the eyelids against the orbital rim; never press on the eye itself.
  3. Aim for thin, even fluorescein rings whose inner edges just touch; too much fluorescein gives a falsely high reading.
  4. With high corneal astigmatism (more than about 3 D), adjust the prism position.
  5. Record the method, the time of day and the corneal thickness with every reading.

Test yourself

Five questions. Choose an answer to see the explanation.

1. Which method is the reference standard for measuring eye pressure?
2. A patient has a central corneal thickness of 480 µm and a Goldmann reading of 19 mmHg. What is most likely?
3. Why can one office reading be misleading?
4. After LASIK for myopia, a patient's Goldmann IOP is 14 mmHg. How should you interpret this?
5. While measuring, you hold the eye open by pressing on the upper eyelid. What is the effect on the reading?