Ahmed Glaucoma Valve
A drainage device with a built-in valve that limits outflow right after surgery.
In short
Ahmed Glaucoma Valve Explained
Plays on YouTube. By pressing play you agree to YouTube's cookies.
Animation and script by Ronald de Crom, ophthalmologist and glaucoma specialist. Narration: synthetic voice.
The Ahmed Glaucoma Valve is a drainage device with a tube and a plate on the back of the eye. A built-in valve limits the outflow, which lowers the risk of a pressure that is too low right after surgery.
It is used for difficult glaucoma, for example after a failed operation or in secondary glaucoma.
In depth
- Device
- A flow-restrictive valve mechanism with silicone membranes; plates in silicone or polypropylene, in several sizes.
- Evidence
- Compared with the non-valved Baerveldt implant (ABC and AVB studies), the Ahmed valve had fewer early hypotony-related complications, but slightly higher IOP and more need for medication in the long term.
- Hypertensive phase
- A rise in IOP in the first weeks to months after surgery is common and is usually managed with drops.
- Complications
- Hypertensive phase, tube obstruction or exposure, endothelial cell loss and double vision.
Surgical technique
Ahmed Glaucoma Valve – Step by Step
Plays on YouTube. By pressing play you agree to YouTube's cookies.
Surgical footage is shown with written patient consent.
Surgical steps
- Prime the valve: flush BSS through the tube with a 27G or 30G cannula until fluid leaves the outlet
- Fornix-based peritomy superotemporally over about 90°, with a relaxing incision if needed
- Sub-Tenon pocket posteriorly between the superior and lateral rectus muscles
- Slide the plate into the pocket, valve housing first, between (not under) the rectus muscles
- Suture the plate with nylon through both eyelets, leading edge 8–10 mm from the limbus
- Clear-corneal paracentesis and viscoelastic in the anterior chamber
- Trim the tube bevel up so the tip will lie 2–3 mm into the anterior chamber
- 23G needle track about 2 mm behind the limbus, parallel to the iris
- Insert the tube into the anterior chamber, bevel towards the cornea
- Suture the tube to the sclera with nylon, without compressing the lumen
- Patch graft (donor sclera, pericardium or cornea) over the tube
- Exchange most of the viscoelastic for BSS
- Close conjunctiva and Tenon at the limbus with 8-0 Vicryl
Pearls
- Always prime the valve: an unprimed valve may not open.
- Avoid placing the plate too far forward, to reduce the risk of exposure.
Also in Drainage implants:
PAUL Glaucoma Implant