PAUL Glaucoma Implant
A drainage device with a very thin tube that carries fluid to a plate on the back of the eye.
In short
PAUL Glaucoma Implant Explained
Plays on YouTube. By pressing play you agree to YouTube's cookies.
Animation and script by Ronald de Crom, ophthalmologist and glaucoma specialist. Narration: synthetic voice.
The PAUL Glaucoma Implant is a drainage device. A thin silicone tube carries fluid from inside the eye to a plate that is stitched onto the wall of the eye, further back. Fluid collects around the plate and is absorbed.
It is used when other treatments have not worked or are unlikely to work, for example after an earlier operation that has failed.
In depth
- Device
- A non-valved device with a large endplate and a small-calibre tube (outer diameter 467 µm, inner diameter 127 µm), designed to take up less space in the anterior chamber.
- Flow control
- As the device has no valve, early hypotony is prevented by temporarily occluding the tube, for example with an intraluminal stent and/or a ligature.
- Indications
- Refractory glaucoma, failed filtering surgery, secondary glaucomas (such as uveitic or neovascular glaucoma) and eyes with conjunctival scarring.
- Complications
- Hypotony, tube obstruction or exposure, double vision and corneal endothelial cell loss.
Surgical technique
PAUL Glaucoma Implant – Step by Step
Plays on YouTube. By pressing play you agree to YouTube's cookies.
Surgical footage is shown with written patient consent.
Surgical steps
- Conjunctival and Tenon's incision about 5 mm behind the limbus in the chosen quadrant
- Generous sub-Tenon pocket posteriorly, between and just under the two rectus muscles, and a smaller pocket towards the limbus
- Mitomycin C on sponges in the posterior pocket; remove them and irrigate generously
- Slide the end plate into the pocket, with both wings under the adjacent rectus muscles
- Suture the plate to the sclera with nylon through both eyelets, leading edge about 10 mm from the limbus
- Side port in clear cornea, carbachol (Miostat) in the anterior chamber, then viscoelastic
- Small scleral pocket 3 mm from the limbus and a 25G needle tract into the anterior chamber, parallel to the iris
- Trim the tube bevel up so the tip will lie 2–3 mm into the anterior chamber
- Insert the tube into the anterior chamber, bevel towards the cornea
- 6-0 Prolene stent in the tube lumen
- Suture the tube to the sclera with nylon, without compressing the lumen
- Donor sclera over the tube, from the limbus to over the tube suture
- Wash out the viscoelastic with BSS
- Lay the end of the stent under the conjunctiva at the limbus
- Close Tenon and conjunctiva with a running Vicryl suture
Pearls
- Keep the tube clear of the corneal endothelium; sulcus placement is an option in pseudophakic eyes.
- Plan when to remove the stent or release the ligature to control the IOP in the first weeks.
Also in Drainage implants:
Ahmed Glaucoma Valve