Animation and script by Ronald de Crom, ophthalmologist and glaucoma specialist. Narration: synthetic voice.
Learning objectives
After this module you can:
- Know the device and its indication
- Perform the key steps and position the tube correctly
- Recognise complications and how they compare with trabeculectomy
Explore step by step
The view through the operating microscope, with the surgeon at the head of the table: cornea at the top, superior fornix at the bottom. Eleven steps, from peritomy to closure, plus an optional step with fixation and stent. Choose a step, or press Play.
Animation and illustration: R. de Crom. Schematic representation; doses and materials according to local protocol.
Drawn to scale (1 mm scale bar, bottom right). PreserFlo: length 8.5 mm, outer diameter 0.35 mm, lumen 70 µm, fin 1.1 mm wide. Proximal segment 4.5 mm (tunnel plus intracameral part), distal tail 3 mm. Scleral pocket 1.0 mm wide and about 1 mm long, at the mark 3 mm behind the limbus; the fin sits entirely within it. The distal end lies about 6 mm behind the limbus.
Key points
The PreserFlo MicroShunt is a tube made of SIBS, 8.5 mm long with a 70 µm lumen, implanted ab externo with mitomycin C. No scleral flap or sclerostomy is needed.
Indication: mild to advanced open-angle glaucoma not controlled on maximal tolerated medical therapy.
In a randomised comparison with trabeculectomy, the IOP reduction was somewhat smaller, with fewer postoperative interventions.
The tube should lie in the anterior chamber parallel to the iris, clear of the cornea, with the fin wedged in the scleral pocket.
Confirm spontaneous flow at the distal end before closure; if there is none, prime the tube with a cannula.
Complications: early hypotony, bleb fibrosis and failure, and (rarely) tube exposure, migration or contact with the cornea or iris.
Test yourself
Five questions. Choose an answer to see the explanation.