PAUL Glaucoma Implant: technique

  • Module 6 of 15
  • About 15 minutes
  • Level 3
PAUL Glaucoma Implant – Step by Step Plays on YouTube. By pressing play you agree to YouTube's cookies.

Animation and script by Ronald de Crom, ophthalmologist and glaucoma specialist. Narration: synthetic voice.

Learning objectives

After this module you can:

  • Know the device and its indications
  • Perform the key steps, including flow restriction with a stent
  • Recognise the main complications

Explore step by step

The view through the operating microscope, with the surgeon at the head of the table: cornea at the top, fornix at the bottom. Fifteen steps, from incision to closure. Choose a step, or press Play.

    Animation and illustration: R. de Crom. Schematic representation; doses and materials according to local protocol.

    Drawn to scale (1 mm scale bar, bottom right). PAUL end plate 21.9 mm wide and 16.1 mm deep (342 mm²), leading edge about 10 mm from the limbus; tube outer diameter 0.467 mm, lumen 0.127 mm. The plate wings lie under the adjacent rectus muscles. Quadrant centred in view.

    Key points

    • The PAUL Glaucoma Implant is a non-valved drainage device with a large end plate (342 mm²) and a small-calibre tube (outer diameter 0.467 mm, lumen 0.127 mm) that takes up little space in the anterior chamber.

    • Indications: refractory glaucoma, failed filtering surgery, secondary glaucomas such as uveitic or neovascular glaucoma, and eyes with conjunctival scarring.

    • Open the conjunctiva about 5 mm behind the limbus, so the wound lies away from the tube and patch. The plate wings go under the adjacent rectus muscles, and the plate is sutured with its leading edge about 10 mm from the limbus. Mitomycin C is applied posteriorly, according to local protocol.

    • The tube enters the anterior chamber through a 25G needle tract from a small scleral pocket 3 mm from the limbus, parallel to the iris. Trim it so the tip lies 2–3 mm into the chamber, bevel towards the cornea, clear of endothelium and iris; err on the short side.

    • Because there is no valve, a 6-0 Prolene stent in the lumen limits early flow and prevents hypotony. Its end is left under the conjunctiva at the limbus, so it can be removed at the slit lamp when more flow is needed. A donor scleral patch covers the tube.

    • Complications: hypotony (often after stent removal), tube obstruction or exposure, double vision and corneal endothelial cell loss.

    Test yourself

    Five questions. Choose an answer to see the explanation.

    1. Why does the PAUL implant need a stent in the tube early after surgery?
    2. Where do the wings of the PAUL end plate lie?
    3. Why is the end of the Prolene stent placed at the limbus?
    4. What is the purpose of the donor scleral patch?
    5. How far should the tube tip extend into the anterior chamber?