Ahmed Glaucoma Valve: technique

  • Module 7 of 15
  • About 15 minutes
  • Level 3
Ahmed Glaucoma Valve – Step by Step Plays on YouTube. By pressing play you agree to YouTube's cookies.

Animation and script by Ronald de Crom, ophthalmologist and glaucoma specialist. Narration: synthetic voice.

Learning objectives

After this module you can:

  • Know how the valve works and why it must be primed
  • Perform the key steps of implantation
  • Recognise the hypertensive phase and the main complications

Explore step by step

The view through the operating microscope, with the surgeon at the head of the table: cornea at the top, fornix at the bottom. First, priming the valve on the table. Thirteen steps, from priming to closure. Choose a step, or press Play.

    Animation and illustration: R. de Crom. Schematic representation; doses and materials according to local protocol.

    Drawn to scale (1 mm scale bar, bottom right). Ahmed FP7: 16 × 13 mm silicone plate (184 mm²) with valve housing and venturi-shaped valve chamber; leading edge 8–10 mm from the limbus, between the superior and lateral rectus (right eye, superotemporal). Tube outer diameter 0.635 mm, lumen 0.305 mm. MMC is optional per local protocol and not shown.

    Key points

    • The Ahmed Glaucoma Valve has a flow-restricting valve of silicone membranes. The FP7 has a 16 × 13 mm silicone plate (184 mm²) and a tube with an outer diameter of 0.635 mm. Because the valve limits early outflow, no ligature or stent is needed.

    • Always prime the valve before implantation: flush BSS through the tube with a 27G or 30G cannula until fluid leaves the outlet. This separates membranes that may stick together; an unprimed valve can stay closed.

    • The superotemporal quadrant gives the most room and the upper lid covers the plate. The plate lies between the superior and lateral rectus muscles, not under them, and is sutured with its leading edge 8–10 mm from the limbus.

    • The tube enters the anterior chamber through a 23G needle track about 2 mm behind the limbus, parallel to the iris. Trim it so the tip lies 2–3 mm into the chamber, bevel towards the cornea, and cover it with a patch graft.

    • A hypertensive phase, a rise in pressure in the first weeks to months, is common and is usually managed with drops.

    • Compared with the non-valved Baerveldt implant (ABC and AVB studies), the Ahmed valve had fewer early hypotony-related complications, but a slightly higher pressure and more medication in the long term. Other complications: tube obstruction or exposure, endothelial cell loss and double vision.

    Test yourself

    Five questions. Choose an answer to see the explanation.

    1. Why must the Ahmed valve be primed before implantation?
    2. Which quadrant is usually chosen for the Ahmed valve, and why?
    3. How is the Ahmed FP7 plate positioned relative to the rectus muscles?
    4. Some weeks after surgery the pressure rises to 24 mmHg with a well-positioned, patent tube. What is the most likely cause?
    5. What did the ABC and AVB studies show for the Ahmed valve compared with the Baerveldt implant?